OBJECTIVE
To summarize the principles and application of phototherapy consistent with the current 2022 American Academy of Pediatrics “Clinical Practice Guideline Revision for the Management of Hyperbilirubinemia in the Newborn Infant 35 or More Weeks of Gestation.”
METHODS
Relevant literature was reviewed regarding phototherapy devices in the United States, specifically those that incorporate blue to blue-green light-emitting diode, fluorescent, halogen, or fiberoptic light sources, and their currently marketed indications.
RESULTS
The efficacy of phototherapy devices varies widely because of nonstandardized use of light sources and configurations and irradiance meters. In summary, the most effective and safest devices have the following characteristics: (1) incorporation of narrow band blue-to-green light-emitting diode lamps (∼460–490 nm wavelength range; 478 nm optimal) that would best overlap the bilirubin absorption spectrum; (2) emission of irradiance of at least 30 µW/cm2/nm (in term infants); and (3) illumination of the exposed maximal body surface area of an infant (35% to 80%). Furthermore, accurate irradiance measurements should be performed using the appropriate irradiance meter calibrated for the wavelength range delivered by the phototherapy device.
CONCLUSIONS
With proper administration of effective phototherapy to an infant without concurrent hemolysis, total serum or plasma bilirubin concentrations will decrease within the first 4 to 6 hours of initiation safely and effectively.